Detcon Model 972-071170-0LCGas Detection SystemOperator's Instruction ManualOctober 1, 1997 Document #1870 Rev 1.0Detcon Model 972-071170-0LC ga
1.0 DESCRIPTIONThe Detcon Model 972-071170-0LC is a 117VAC powered gas detection system that provides a remote gassample to two independent sensor as
2. Install a length of tubing from the desired sample point to the sample port as shown below. Sample draw tub-ing should be either teflon or stainle
1.2 START UPUpon completion of all tubing connections and field wiring turn on the power switch located on the enclo-sure door. Note that the power l
activated. Typically, air flow should range between 500 and 1000 ml/min. (milliliters per minute) as seenon the flow meter located on the left of the
To Calibrate: (refer to illustration on the previous page)1. Turn the 3-way valve so that the arrow is pointing down. Sample is now being drawn from
1.7 FLOW DIAGRAM1.8 WIRING DIAGRAMShipping address: 3200 Research Forest Dr. A-1, The Woodlands, Texas 77381❏ Phone: (281) 367-4100 ❏ Fax: (281) 298